ABB TTF300.E1C2OPT Field Mounted Temperature Transmitters TTF300-E1C2/OPT Hart Transmitters In Stock
1. General Introduction
ABB TTF300 is a field mount temperature transmitter with SIL approval. It accepts dual sensors input of RTD, thermocouple, Ohm and mV and outputs 4 to 20 mA HART, PROFIBUS PA or FOUNDATION Fieldbus. Accuracy is 0.1 % with long-term stability 0.05 % per year, extended diagnostics in accordance with NE 107,IP 66 / IP 67, NEMA 4X aluminum or stainless steel field-mount housing optional.
2. Measuring Span & Accuracy
Sensor | Measuring range limits | Minimum span | Input (24-bit AD-converter) | Output (16-bit DA) | |
Resistance thermometer/resistor | |||||
DIN IEC 60751 | Pt10 (a=0.003850) | -200 ... 850 °C (-328 … 1562 °F) | 10 °C (18 °F) | ± 0.80 °C (± 1.44 °F) | ± 0.05 % |
Pt50 (a=0.003850) | ± 0.16 °C (± 0.29 °F) | ± 0.05 % | |||
Pt100 (a=0.003850 | ± 0.08 °C (± 0.14 °F) | ± 0.05 % | |||
Pt200 (a=0.003850) | ± 0.24 °C (± 0.43 °F) | ± 0.05 % | |||
Pt500 (a=0.003850) | ± 0.16 °C (± 0.29 °F) | ± 0.05 % | |||
Pt1000 (a=0.003850) | ± 0.08 °C (± 0.14 °F) | ± 0.05 % | |||
JIS C1604 | Pt10 (a=0.003916) | -200 ... 645 °C (-328 ... 1193 °F) | 10 °C (18 °F) | ± 0.80 °C (± 1.44 °F) | ± 0.05 % |
Pt50 (a=0.003916) | ± 0.16 °C (± 0.29 °F) | ± 0.05 % | |||
Pt100 (a=0.003916) | ± 0.08 °C (± 0.14 °F) | ± 0.05 % | |||
MIL-T-24388 | Pt10 (a=0.003920 | -200 ... 850 °C (-328 ... 1562 °F) | 10 °C (18 °F) | ± 0.80 °C (± 1.44 °F) | ± 0.05 % |
Pt50 (a=0.003920) | ± 0.16 °C (± 0.29 °F) | ± 0.05 % | |||
Pt100 (a=0.003920) | ± 0.08 °C (± 0.14 °F) | ± 0.05 % | |||
Pt200 (a=0.003920) | ± 0.24 °C (± 0.43 °F) | ± 0.05 % | |||
Pt1000 (a=0.003920) | ± 0.08 °C (± 0.14 °F) | ± 0.05 % | |||
DIN 43760 | Ni50 (a=0.006180) | -60 ... 250 °C (-76 ... 482 °F) | 10 °C (18 °F) | ± 0.16 °C (± 0.29 °F) | ± 0.05 % |
Ni100 (a=0.006180) |
± 0.08 °C (± 0.14 °F) | ± 0.05 % | |||
Ni120 (a=0.006180) | ± 0.05 % | ||||
Ni1000 (a=0.006180) | ± 0.05 % | ||||
OIML R 84 | Cu10 (a=0.004270) | -50 ... 200 °C (-58 ... 392 °F) | 10 °C (18 °F) | ± 0.80 °C (± 1.44 °F) | ± 0.05 % |
Cu100 (a=0.004270) | ± 0.08 °C (± 0.14 °F) | ± 0.05 % | |||
Resistance measurement | 0 ... 500 Ω | 4 Ω | ± 32 mΩ | ± 0.05 % | |
0 … 5000 Ω | 40 Ω | ± 320 mΩ | ± 0.05 % | ||
Thermocouples/ voltages | |||||
IEC 60584 | Type K (Ni10Cr-Ni5) | -270 … 1372 °C (-454 … 2502 °F) | 50 °C (90 °F) |
± 0.35 °C (± 0.63 °F) | ± 0.05 % |
Type J (Fe-Cu45Ni) | -210 … 1200 °C (-346 … 2192 °F) | ± 0.05 % | |||
Type N (Ni14CrSi-NiSi) | -270 … 1300 °C (-454 … 2372 °F) | ± 0.05 % | |||
Type T (Cu-Cu45Ni) | -270 ... 400 °C (-454 ... 752 °F) | ± 0.05 % | |||
Type E (Ni10Cr-Cu45Ni) | -270 … 1000 °C (-454 … 1832 °F) | ± 0.05 % | |||
Type R (Pt13Rh-Pt) | -50 … 1768 °C (-58 … 3215 °F) | 100 °C (180 °F) |
± 0.95 °C (± 1.71 °F) | ± 0.05 % | |
Type S (Pt10Rh-Pt) | ± 0.05 % | ||||
Type B (Pt30Rh-Pt6Rh) | -0 … 1820 °C (32 … 3308 °F) | ± 0.05 % | |||
DIN 43710 | Type L (Fe-CuNi) | -200 … 900 °C (-328 … 1652 °F) | 50 °C (90 °F) | ± 0.35 °C (± 0.63 °F) | ± 0.05 % |
Type U (Cu-CuNi) | -200 … 600 °C (-328 … 1112 °F) | ± 0.05 % | |||
ASTM E 988 | Type C | -0 … 2315 °C (32 … 4200 °F) | 100 °C (180 °F) | ± 1.35 °C (± 2.43 °F) | ± 0.05 % |
Type D | ± 0.05 % | ||||
Voltage measurement | -125 ... 125 mV | 2 mV | ± 12 μV | ± 0.05 % | |
-125 ... 1100 mV | 20 mV | ± 120 μV | ± 0.05 % |
3. Model selection
Model | Suffix Codes | Description | ||||
TTF300 | TTF300 Field Mounted Temperature Transmitter | |||||
Explosion Protection | Y0 | Without explosion protection | ||||
E1 | ATEX Intrinsic Safety | |||||
E3 | ATEX Flameproof | |||||
E4 | ATEX Flameproof and Intrinsic Safety | |||||
H1 | IECEx Intrinsic Safety | |||||
H5 | IECEx Flameproof type of protection | |||||
L1 | FM Intrinsic Safety | |||||
L2 | FM Non-incendive | |||||
L3 | FM Explosion-proof | |||||
S1 | NEPSI Ex ia IIC T6 | |||||
Housing / Display | A | Single-compartment housing (aluminium) / Without display | ||||
B | Single-compartment housing (stainless steel) / Without display | |||||
C | Single-compartment housing (aluminium) / With LCD-display HMI | |||||
D | Single-compartment housing (stainless steel) / With LCD-display HMI | |||||
Cable Entry | 1 | Thread 2 × M20 × 1.5 | ||||
2 | Thread 2 × 1/2 in NPT | |||||
3 | Thread 2 × 3/4 in NPT | |||||
4 | Cable gland 2 x M20 × 1.5 | |||||
Communication protocol | H | HART®, programmable, output signal 4 to 20 mA, dual input | ||||
P | PROFIBUS PA | |||||
F | FOUNDATION fieldbus | |||||
Configuration | BS | Standard configuration) | ||||
BF | Customer-specific configuration, except user curve | |||||
BG | Customer-specific configuration, including user curve | |||||
Optional codes | /Optional codes |